Numerical Comparison between MSRC and AC based on a Radiated Susceptibility Test
Keywords:
Mode Stirred Reverberation Chamber (MSRC)Anechoic Chamber (AC)Electromagnetic Compatibility (EMC)Abstract
This study involves a comparison of the experimental findings obtained from testing conducted in the Mode Stirred Reverberation Chamber (MSRC) and the Anechoic Chamber (AC). Directly comparing the reactions of different items under test proved challenging due to variations in the electromagnetic surroundings for both procedures. The tests conducted in both rooms have exhibited varying responses based on the equipment's directivity. Furthermore, the outcomes derived from this examination exhibit variability contingent upon the conditions under which the test is conducted. Hence, the test results obtained from the two chambers exhibit similar error biases. The error bias refers to the proportion of a measured response obtained under specified test conditions compared to the maximum possible reaction. The paper examines the coupling uncertainty and anticipated error bias for both test procedures, analyzing how they vary with apparent directivity. The measured AC data is utilized to ascertain the magnitude and configuration of the apparent directivity of equipment responses.